  • Outstanding tear strength and elongation. Strong, flexible impression material stretches around undercuts without tearing or distorting.
  • Hydrophilic dental wash material. Displaces intraoral fluids to produce impressions with great detail and no voids or gaps.
  • Excellent dimensional stability. Impression will not distort due to dimensional change or temporary deformation.
  • Widest selection of viscosities, set times and delivery methods. Helps overcome common difficulties in impression taking, and matches doctor preference.

Take 1 Advanced vinylpolysiloxane (VPS) impression material employs Kerr's long history of expertise in developing high-performance impression materials, along with recent technological advancements, to deliver a VPS impression material with the optimum combination of physical properties: strength, elasticity, dimensional stability and the ability to register detail in any environment. Furthermore, Take 1 Advanced comes in the widest selection of set times, delivery systems and consistencies available on the market today - so you always have the perfect combination of mechanical properties in a material that feels like it was designed just for your technique preferences.

Kits

34064Take 1 Advanced- Cartridge Intro Kit2 Cartridges (50ml) LB Wash 3 Cartridges (50 ml) Tray (Fast Set) 1 Cartridge (50 ml) Bite Registration 1 Extruder 6 Small Mix tips 6 Intraoral Syringe Tips (Small Universal) 6 Large Mix tips

Two Packs

33951Take 1 Advanced Two-Pack - Light Body Wash - Regular Set (Orange)2 Cartridges (50ml), 6 Mixing Tips
33953Take 1 Advanced Two-Pack - Light Body Wash - Fast Set (Orange)2 Cartridges (50ml), 6 Mixing Tips
33955Take 1 Advanced Two-Pack - Light Body Wash - Super Fast Set (Orange)2 Cartridges (50ml), 6 Mixing Tips
33957Take 1 Advanced Two-Pack - Medium Body/Monophase Regular Set (Light Blue)2 Cartridges (50ml), 6 Mixing Tips
33959Take 1 Advanced Two-Pack - Medium Body/Monophase - Fast Set (Light Blue)2 Cartridges (50ml), 6 Mixing Tips
33961Take 1 Advanced Two-Pack - Medium Body/Monophase - Super Fast (Light Blue)2 Cartridges (50ml), 6 Mixing Tips
33963Take 1 Advanced Two-Pack - Tray - Regular Set (Dark Blue)2 Cartridges (50ml), 6 Mixing Tips
33965Take 1 Advanced Two-Pack - Tray - Fast Set (Dark Blue)2 Cartridges (50ml), 6 Mixing Tips
33967Take 1 Advanced Two-Pack - Tray - Super Fast Set (Dark Blue)2 Cartridges (50ml), 6 Mixing Tips
33975Take 1 Advanced Two-Pack - Bite Registration2 Cartridges (50ml), 6 Mixing Tips
34457Take 1 Advanced Two-Pack - Rigid Tray - Fast Set (Dark Green)2 Cartridges (50ml), 6 Mixing Tips

Value Packs

33952Take 1 Advanced Value Pack - Light Body Wash - Regular Set (Orange)24 Cartridges (50 ml)
33954Take 1 Advanced Value Pack - Light Body Wash - Fast Set (Orange)24 Cartridges (50 ml)
33956Take 1 Advanced Value Pack - Light Body Wash - Super Fast Set (Orange)24 Cartridges (50 ml)
33958Take 1 Advanced Value Pack - Medium Body/Monophase - Regular Set (Light Blue)24 Cartridges (50 ml)
33960Take 1 Advanced Value Pack - Medium Body/Monophase - Fast Set (Light Blue)24 Cartridges (50 ml)
33962Take 1 Advanced Value Pack - Medium Body/Monophase - Super Fast (Light Blue)24 Cartridges (50 ml)
33964Take 1 Advanced Value Pack - Tray - Regular Set (Dark Blue)24 Cartridges (50 ml)
33966Take 1 Advanced Value Pack - Tray - Fast Set (Dark Blue)24 Cartridges (50 ml)
35186Take 1 Advanced Value Pack - Rigid Tray - Fast Set (Dark Green)24 Cartridges (50 ml)
35187Take 1 Advanced Value Pack - Rigid Tray - Super Fast Set (Dark Green)24 Cartridges (50ml)

Accessories

16622VPS Adhesive15 ml bottle
20731Intraoral Syringe Tips, Large 100 disposables
25777VPS Adhesive 60 ml bottle
27517Intraoral Syringe Tips, Small Universal100 disposables
29333Mix Tip Refill - Largepkg of 48 tips
29363Mix Tips Refill - Smallpkg of 48 tips
34746Dynamic Mixing Tips for Volume 50 tips, 2 bayonet rings

Volume

34688Medium/Monophase Regular Set2 Hard Cartridges (380 ml)
34689Medium/Monophase Fast Set2 Hard Cartridges (380 ml)
34690Medium/Monophase Super Fast Set2 Hard Cartridges (380 ml)
34695Volume Tray Regular Set2 Hard Cartridges (380 ml)
34696Volume Tray Fast Set2 Hard Cartridges (380 ml)
34697Volume Tray Super Fast Set2 Hard Cartridges (380 ml)
34825Rigid Tray Fast Set (Dark Green)2 Hard Cartridges (380 ml)
34826Rigid Tray Super Fast Set (Dark Green)2 Hard Cartridges (380 ml)

Putty

34070Putty1 Base (400 g), 1 Catalyst (400 g)

Troubleshooting Guide

Problem

Evidence Cause Solution

Premature Set

impression_illo10-100x100

  • Abrupt change from heavy to light-body material
  • Muted detail around margin
  • Set time of wash/tray material not synched
  • Exceeded working time of material
  • Begin mix of heavy and light viscosity concurrently
  • Verify set times for heavy and light viscosities match
  • Observe working time of materials

Deformation/Dimensional Change

impression_illo4-100x100-2

  • Crowns too tall/short
  • Open contacts
  • Crown will not seat
  • Insufficient mix of material
  • Impression removed too early
  • Tray movement after seating
  • Insufficient material elasticity
  • Fluid absorption during disinfection process
  • Verify material is adequately mixed before seating and completely set prior to removal
  • Use passive force to maintain tray position when using open-bite technique
  • Verify no tooth contact with tray sides or impingement in closed-bite technique
  • Choose material with adequate elasticity
  • Follow disinfection protocol carefully

Tearing at Margin

impression_illo1-100x100

  • Visible tears at margin
  • Lack of flash extending apically around entire preparation
  • Material has insufficient tear strength
  • Presence of severe undercuts
  • Insufficient sulcular expansion
  • Use material with adequate tear strength
  • Consider blocking out undercuts, especially in cases with gingival recession
  • Ensure at least .5 mm sulcular expansion

Surface Inhibition or “Non-set”

impression_illo2.-100x100jp

  • Unset or mottled surface around preparation
  • Lack of detail
  • Shiny/wet appearance
  • Contamination due to latex gloves; direct contact with material or residue left on teeth
  • Contamination due to temporary materials or composite
  • Expired material or exceeded material shelf life
  • Use latex-free gloves when handling VPS material
  • Rinse preparation area thoroughly after using other restorative materials
  • Use automated mixing device to minimize exposure to contaminants
  • Do not use expired material

Poor model detail

impression_illo3-100x100

  • Small bubbles/indentations in cast
  • Powdery cusps
  • Release of hydrogen gas from VPS impression after pouring cast
  • Tooth contact with mesh lining of closed-bite impression tray
  • Follow manufacturer instructions regarding any delays for pouring models after taking impression
  • Place cotton roll on contra-lateral side when using closed-bite trays to prevent direct tooth contact with tray; this prevents water from leaching through tray after cast has been poured

Voids in wash material

impression_illo5-100x100

  • Visible voids in wash material around tooth preparation
  • Bubbles incorporated into material during mixing or loading syringe
  • Bubbles incorporated while syringing material intraorally
  • Use automix systems
  • Syringe material intraorally in continuous stream around each preparation; do not lift syringe tip from material

Insufficient or excessive compression

impression_illo10-100x100

  • Insufficient: wash material flows away from preparation area
  • Excessive: wash material displaced from preparation area; “burn through”
  • Insufficient compression of wash material from adjacent tooth/tray or heavy-body material
  • Excessive compression of wash material due to high viscosity of tray material
  • Use tray viscosity material with greater compressive force, or use custom tray
  • Use tray material with lower viscosity to avoid displacing wash material

Incomplete margin

impression_illo8-100x100

  • Voids or “jumps” on margin
  • Inadequate tissue management
  • Contamination from intraoral fluids
  • Insufficient compression of wash material
  • Use Expasyl™ retraction paste to open sulcus and help maintain dry field
  • Use hydrophilic wash material
  • Select tray viscosity with adequate compression

Tooth contact with trayimpression_illo10-100x100

  • Visible impingement
  • Incorrect size tray
  • Tray seating not aligned with dentition
  • Verify tray has enough room for 2 mm of impression material between tooth and tray walls

Documentation

Document types

Language

Sort by

Elements per page

Displaying 1 - 7 of 7
Kerr Restorative Catalog
Document Type: Product Brochure
Language: English (US)
Release date: 
Wednesday, June 15, 2022 - 02:06
File size: 
2.593
Take 1 Advanced - Sales Sheet
Document Type: Product Brochure
Language: English (US)
Release date: 
Friday, July 24, 2015 - 01:07
File size: 
0.487
TF Twisted Files
Document Type: Product Brochure
Language: English (US)
Release date: 
Monday, July 27, 2015 - 01:07
File size: 
0.644
TAKE 1 ADVANCED BITE REGISTRATION eIFU_NA
Document Type: Instructions for Use (IFU)
Language: English (US), French (Canada)
Release date: 
Friday, May 12, 2023 - 02:05
File size: 
0.154
TAKE 1 ADVANCED eIFU_NA
Document Type: Instructions for Use (IFU)
Language: English (US), French (Canada)
Release date: 
Friday, May 12, 2023 - 01:05
File size: 
0.175
VOLUME CARTRIDGE TECHNIQUE CARD
Document Type: Instructions for Use (IFU)
Language: English (US)
Release date: 
Monday, February 27, 2023 - 01:02
File size: 
1.642
VPS TRAY ADHESIVE eIFU_NA
Document Type: Instructions for Use (IFU)
Language: English (US), French (Canada)
Release date: 
Monday, February 27, 2023 - 01:02
File size: 
0.14